Product Summary |
|
|
An easy-to-use constant temperature water tank with an integrated shaking mechanism. Ideal not only for culturing but also for reaction and lysis |
|
|
|
|
Features |
|
|
A high-temperature distribution of ± 0.1 ° C can be obtained in the tank due to the stirring function of the pump. It can be used as an accurate water tank even when shaking is not used. |
|
The shaking width can be changed from 10 to 40 mm. The number of shakes is also a step-less speed change of 20 to 160 rpm. |
|
Equipped with safety functions such as automatic overheat prevention, temperature sensor abnormality, and overcurrent leakage breaker as standard equipment. |
|
The top lid enhances heat retention and prevents water from evaporating, making it convenient for long-term use. External communication, temperature output, time-up output, and external alarm are set, and 4 options can be selected at the same time. |
|
|
|
|
|
|
|
Specifications |
|
|
Specification Item |
Details |
|
Model |
BT101 |
|
Method |
Reciprocating shaking system, Stirring by pump |
|
Operating temperature range |
Room temp.+5°C~80°C(when top cover is installed) |
|
Temperature control accuracy |
±0.02°C to 0.08°C |
|
Temperature distribution accuracy |
±0.1°C |
|
Temperature rising time |
Approx. 95 min. |
|
Shaking frequency |
20 to 160 times/min |
|
Shaking width |
10 to 40mm (adjustable) |
|
Material (Interior) |
Stainless steel |
|
Temp. Controller |
PID control by microprocessor |
|
Temp. Sensor |
Platinum resistance temperature detector Pt100Ω (for temperature control) / K thermocouple (for overheating inhibitor) |
|
Temp setting/display method |
Digital setting |
|
Overheat protector |
ON/OFF control by microprocessor |
|
Overheat protect setting method |
Digital setting |
|
Heater |
Copper pipe heater (nickel plated) 1.2kW |
|
Stirring System |
Magnet pump 6W |
|
Shaker |
Gear motor 25W |
|
Shaking speed setting system |
Analog setting |
|
Shaking speed display system |
Digital display |
|
Timer |
1 min. to 99 hr. 59 min. Digital display, Quick automatic start/stop |
|
Safety countermeasures |
Self-diagnosis function(Heater defective, Sensor defective, SSR short circuit and Overheat protector) |
|
Safety countermeasures |
over current, short circuit breaker, Heater no-load operation prevention device and overheat protector. |
|
Container flame number, Test tube |
Dia : 16.5mm, Length : 150 to 200mm(JIS) 130 pcs. |
|
Flask |
Round, Erlenmeyer flask 100/300/500ml, 12/5/3pcs. |
|
Shaking Rack (Dimensions W×D×Hmm) |
1 pc. 290×220×170 (268×200×150) |
|
Spring Pitch |
20mm (both length and width) |
|
Internal dimensions(W×D×H mm) |
302×350×250mm |
|
External dimensions(W×D×H mm) |
579×414×325mm |
|
Remarks(dimensions) |
19L(When the water is full) |
|
Power Source |
Single phase AC100V14A |
|
Weight |
27kg |