Two supports are fixed onto the base with nuts, and their upper ends are secured to the upper beam. Two additional supports are fixed to the middle beam and operating platform, together forming a space for fixing specimens during tension, compression, and bending tests.
When oil from the pump lifts the piston, the operating platform rises, raising the two supports and the upper beam. The upper jaw and middle jaw are located on the upper beam and middle beam, respectively. The middle beam is raised or lowered by screws running on the two supports. A lift control button with up/down markings is provided.
The oil cylinder and piston are the most critical components of the main body. Their contacting surfaces are precision-finished, with appropriate clearance and oil film, allowing the piston to move freely and minimizing friction. As oil flows from the pump into the cylinder, the piston rises along with the beam and operating platform until load is applied to the specimen.
Important: Oil must be kept pure and free from impurities or iron dust. Contaminants entering the cylinder via the pump or valves can damage the surface finish of the cylinder and piston, affecting test accuracy. During installation, any dirt or soil on the upper contacting surfaces of the cylinder and piston must be cleaned first.
The machine is supplied with various tension and compression accessories.
The dynamometer consists of a digital display controller, sensor, high-pressure oil pump, and other operating components. All parts are enclosed in a simple, smooth, and aesthetically designed steel shell. Each shell surface has a door for inspection, adjustment, and maintenance, while also keeping bearings and precision parts clean and in good working condition.
The high-pressure oil pump and motor are flange-connected to the oil tank cover. The pump is an axial plunger type consisting of seven sets of pistons. The pistons and sleeves have high surface finish and precise fit, ensuring high pressure capability and minimal oil leakage.
The high-pressure pump motor button, power button, and indicator lights are all installed on the dynamometer panel. The feed oil valve (also known as the pressure regulating valve) controls oil flow from the pump to the cylinder and regulates loading speed. The reflux valve unloads pressure, allowing oil in the working cylinder to return to the tank.
Oil from the tank passes through a strainer into the pump, then is delivered to the feed valve via the pump pipeline.
When the feed oil handwheel is closed: Piston is pushed open by oil pressure, and oil returns to the tank.
When the feed oil handwheel is open: Oil enters the working cylinder through the pipeline, then passes through the pressure pipeline and reflux valve body before returning to the tank.
| Model | EI-XLC300B | EI-XLC600D | EI-XLC1000D / EI-XLC1200D | EI-XLC2000D | |
| Structure | Four Column with Two Reeling Screws Load Frame | ||||
| Maximum Load Capacity | 300kN | 600kN | 1000kN / 1200kN | 2000kN | |
| Force Range | 2% ~ 100% FS | ||||
| Force Accuracy | ±1% | ||||
| Resolution | 1/500000 FS | ||||
| Adjustment Range of Stress Rate Control | 1MPa/s ~ 60MPa/s | ||||
| Deformation Accuracy | ±1% | ||||
| Deformation Resolution | 1/500000 of Max Extension | ||||
| Adjustment Range of Strain Rate Control | 0.00025/s ~ 0.0025/s | ||||
| Accuracy of Strain Rate Control | ±2% | ||||
| Displacement Resolution | 0.001mm | ||||
| Adjustment Range of Displacement Rate Control | 0.00025/s ~ 0.0025/s | ||||
| Accuracy of Displacement Rate Control | ±2% | ||||
| Range of Holding Control of Stress, Strain and Displacement | 0.3% ~ 100%FS | ||||
| Accuracy of Holding Control of Stress | ±1% | ||||
| Tensile Space | 650mm | 650mm | 700mm | 1300mm | |
| Compression Space | 550mm | 550mm | 550mm | 1000mm | |
| Column Space | 430mm | 450mm | 550mm | 450mm | |
| Round Specimen Clamping Range | Ø6 ~ Ø26mm | Ø6 ~ Ø40mm | Ø6 ~ Ø45mm | Ø20 ~ Ø65mm | |
| Flat Specimen Clamping Thickness | 0 ~ 30mm | 0 ~ 50mm | |||
| Maximum Width of Flat Specimen Clamping | 75mm | 140mm | |||
| Pressure Plate Dimension | Ø160mm | Ø300mm | |||
| Rate Voltage | Three Phase, AC415V 50Hz | ||||
| Frame Dimension (mm) | 840 x 700 x 2200 | 910 x 830 x 2300 | 1200 x 880 x 3300 | ||
| Oil Source Console Dimension | 1150 x 800 x 1000mm | ||||
| Weight (Approx.) | 2000kg | 2300kg | 3200kg | 9000kg | |
Malaysia